Malik Muhammad Adil ( 4th year Medical Student Shifa College of Medicine, Islamabad. )
Ali Yawar Alam ( Department of Community Health Sciences*, Shifa College of Medicine, Islamabad. )
Abstract
Objective: To examine the temperature regulation and standardization practices of clinical laboratories in Karachi.
Methods: Forty five clinical laboratories in Karachi were examined for observing the standard protocols for running a lab with particular reference to temperature regulation. A questionnaire to the effect was filled.
Results: Among the 45 labs included, the mean complete blood count performed per day was 52 ± 47. Only 5 (11%) labs had a temperature reader. Thirty (66.7%) labs had an air-conditioner installed, of which only 24 were found in working condition. Maintenance of instruments was carried out every 67.5±.30.6 days. The mean number of haemolyzed samples was 2.3 ± 1.7. Control was available in 24(53.3%) labs, which was used daily in only 10 labs. Quantity of blood was the same in all the tubes in only 33(73.3%) labs.
Conclusion: Commercial laboratories should be properly registered and their quality standardized (JPMA 55:88;2005).
Introduction
Due to lack of registration and absence of checks for quality standardization, test results from laboratories could be under question. This might have further implications on the patients getting treatment as well as the doctors using those results for diagnosis or further management of patients. With an objective to determine the standardization of results, 45 busy laboratories in Karachi were included in the survey. Temperature control, frequency of equipment maintenance and presence or absence of standardized diagnostic techniques were noted.
Methods
The study was a cross sectional survey carried out between April and August 2002. Informed consent was obtained from the laboratory incharge. Forty five busy laboratories of Karachi were included, of these 26 were hospital based, 17 commercial and 2 charity laboratories.
Data was collected through a questionnaire. The laboratories were visited during working hours and practices were observed. Data was analyzed by SPSS Version 10.
Results
In the 45 labs observed, the mean number of complete blood counts performed daily was 52 ± 47. Only 5 (11%) labs had a temperature reader. Thirty (66.7%) labs had an air-conditioner installed. Out of these 30, only 24 were found to be in working condition. Maintenance of instruments was carried out every 67.5± 30.6 days. Number of haemolyzed samples was 2.3 ± 1.7. Control was available in 24(53.3%) labs, but this was used daily in only 10 labs. Quantity of blood was the same in all the tubes in only 33 (73.3%) labs.
Room temperature in these labs was 29.6±2.6o centigrade, whereas the environmental temperature was 34.4±0.8o centigrade. Alternative power supply such as USP or generator was available in 34 (85.6%) labs only.
Discussion
The purpose of this study was to examine the temperature regulation and standardization practices of clinical laboratories in Karachi. The study suggests that standard operating procedures were not being followed in the major clinical laboratories in Karachi. The most obvious fact being presence of temperature reader in only 5 (11%) labs. Alternative power supply was available in 34 (85.6%) labs which could greatly alter the composition of the chemicals and blood samples under analysis giving spurious results. Thirty (66.7%) labs had an air-conditioner installed, of which only 24 were found in working condition. Lack of temperature control could effect results of the tests, as also
| Table. Statistics related to lab operation and practices (n=45). |
|
| | Mean±SD | Range |
|
| No. of CBC per day | 52±47 | 5-200
|
| Room temperature (centigrade) | 29.6±2.6 | 24-34 |
| Environment temperature (centigrade) | 34.4±0.81 | 33.4 |
| Maintenance of instruments (days) | 67.5± 30.6 | 15-90 |
| No. of haemolyzed samples | 2.3± 1.7 | 1-8 |
| Nature of lab | | |
| | No. | % |
| Hospital lab | 26 | 57.8 |
| Commercial lab | 17 | 37.8 |
| Charitable lab | 2 | 4.4 |
| Temperature reader | | |
| Yes | 5 | 11.1 |
| No | 40 | 89.1 |
| Airconditioner | | |
| Yes | 30 | 66.7 |
| No | 15 | 33.3 |
| Aircondition functioning (out of 30) | | |
| Yes | 24 | 80 |
| No | 6 | 20 |
| Test performed | | |
| Manual | 22 | 48.9 |
| Automated instrument | 23 | 51.1 |
| Alternate power supply | | |
| USP | 10 | 22.2 |
| Generator | 24 | 53.3 |
| None | 11 | 24.4 |
| Type of CBC bottle | | |
| Commercially prepared | 13 | 28.9 |
| Prepared in lab | 32 | 71.1 |
| Domestic refrigerator | | |
| Yes | 43 | 95.6 |
| No | 2 | 4.4 |
| Control available | | |
| Yes | 24 | 53.3 |
| No | 21 | 46.7 |
| Control use (out of 24) | | |
| Daily | 10 | 41.7 |
| When problem arises | 14 | 58.3 |
| Quantity of blood in CBC tubes | | |
| Same in all tubes | 33 | 73.3 |
| Variable | 12 | 26.7 |
| Sample withdrawn in tubes | | |
| With needle | 2 | 4.4 |
| Without needle | 43 | 95.6 |
| Co-operation of lab staff | | |
| Freely allowed | 31 | 68.9 |
| Hesitant | 12 | 26.7 |
| Not allowed | 2 | 4.4 |
|
depicted by the number of haemolyzed samples seen during the survey. Similarly, routine use of controls was only done in 41.7% labs.
Occupational safety standards of these labs have already been shown to be low.1 Gloves were used in only 4.5% labs and needle cutter in 64% labs. Previous studies2 have revealed that among infection control practices, hand washing before blood collection from the patient was the least popular ritual with workers from only one private sector labs practicing it.
Commercial laboratories should be properly registered and their quality standardized. The results of this study warrant tight regulatory mechanisms and frequent checks for proper running of these labs and checks for maintenance of the lab equipments. The results of these labs should be frequently checked and standardized with reference labs.
References
1. Mujeeb SA, Adil MM, Altaf A, et al. Infection control practices in clinical laboratories in Pakistan. Infec Contrl Hosp Epidemiol 2003;24:141-2.
2. Tarar MR, Hamid MF. A comparison of infection, control practices in pathology laboratories of Government teaching hospitals and private sector in Lahore. Ann King Edward Medcal Coll 2002;8:244-6.